Abstract

A simple procedure is described to provide an absolute calibration of ultrasonic compressional wave transducers in a water tank. The calibration makes use of the ionic vibration potential by measuring the electric potential wave that propagates with an acoustic wave in an ionic solution. The method has good spatial resolution and has been used to measure the beam profile and amplitude in absolute units of an ultrasonic pulse in a water tank.