Abstract

We designed a photo-ECMO device to speed up the rate of carbon monoxide (CO) removal by using visible light to dissociate CO from hemoglobin (Hb). Using computational fluid dynamics, fillets of different radii (5 cm and 10 cm) were applied to the square shape of a photo-ECMO device to reduce stagnant blood flow regions and increase the treated blood volume while being constrained by full light penetration. The blood flow at different flow rates and the thermal load imposed by forty external light sources at 623 nm were modeled using the Navier-Stokes and convection-diffusion equations. The particle residence times were also analyzed to determine the time the blood remained in the device. There was a reduction in the blood flow stagnation as the fillet radii increased. The maximum temperature change for all the geometries was below 4 °C. The optimized device with a fillet radius of 5 cm and a blood priming volume of up to 208 cm3 should decrease the time needed to treat CO poisoning without exceeding the critical threshold for protein denaturation. This technology has the potential to decrease the time for CO removal when treating patients with CO poisoning and pulmonary gas exchange inhibition.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.