Abstract

Background Endovascular aortic aneurysm repair (EVAR) with stent-grafts is used widely for the treatment of thoracic aortic aneurysms (TAA). Inappropriate design of stent-grafts may lead to complications such as endoleak, stent-graft migration and new entries, causes of which may be inappropriate radial support force or insufficient longitudinal flexibility of the stent-grafts. Material and methods To improve the mechanical performance of the stent-grafts, a type of non-equal-strut stent hoops was proposed, and the influence of structural parameters on the mechanical performance was studied. Results Results of numerical simulation and physical experiments show that by using the proposed non-equal-strut stent hoops, radial support force and longitudinal flexibility of stent-grafts can be reconciled and balanced. Conclusion Results of this study could be used to facilitate radial force control and longitudinal flexibility enhancement in the design of aortic stent-grafts.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.