Abstract
In the implanted biomedical devices, the silicon chips with monopolar stimulation design have been widely applied. To protect the negative-voltage pins of the implanted silicon chip from the electrostatic discharge (ESD) damage, the ESD protection circuit should be carefully designed to avoid any wrong current path under normal circuit operation with the negative voltage. In this article, a new power-rail ESD clamp circuit for the application with an operating voltage of −6 V has been proposed and verified in a 0.18- $\mu \text{m}$ 3.3-V CMOS process. The proposed circuit, realized with only 3.3-V nMOS/pMOS devices, is able to prevent the gate-oxide reliability issue under this −6-V application. With the proposed ESD detection circuit, the turn-on speed of the main ESD clamp device, which is a stacked-nMOS (STnMOS), can be greatly enhanced. The STnMOS with a width of $400~\mu \text{m}$ can sustain over 8-kV human body model (HBM) ESD stress and perform low standby leakage current of ~5.4 nA at room temperature under the circuit operating condition with −6-V supply voltage.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.