Abstract

Instruments for measuring pulse rate, oxygen saturation, and body temperature for Covid-19 patients have been designed using the MAX30100 sensor and the Internet of Things (IoT)-based MLX90614 sensor. The MAX30100 sensor is used to measure pulse rate and oxygen saturation. A non-contact MLX90614 sensor is used to monitor body temperature, with an ultrasonic sensor used to set the maximum distance between the sensor and the object. The measurement results were transferred to the database via the ESP8266 MCU node's Wi-Fi communication line. The stored data can be accessed via a web browser. Compared to the oximeter, the MAX30100 sensor has an average error rate of 1.027% for pulse measurement and 0% for oxygen saturation. The MLX90614 sensor has a 0.42% average error rate when it was compared to the thermo-gun.Thus, the measuring device can function properly and is feasible to use. Furthermore, because there is no direct contact between the human body and the sensor, the instrument can prevent Covid-19 transmission.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.