Abstract
Dynamic lung sound signal analysis and processing require synchronous multi-channel data acquisition system. In noninvasive and no pain lung sound diagnosis system, the highly accurate, synchronous and stable data acquisition system is the foundation of fellow-up work. In order to acquire the array data of lung sound, higher speed, reliability and synchronous data acquisition system can meet the requests. Lung sound signal was produced by lung vibration during air exchange. Simultaneous sampling maintains the phase information and provides us with more detailed information about physiology and pathology. Our method is to use piezopolymer sensor array attached to the backside and capture the lung vibration information. This paper mainly presents a FPGA-based synchronous data acquisition system. FPGA as a main component offers the ADC chip timing and FIFO design using Verilog HDL source code. Furthermore, this paper solves the synchronization problems so that array phase information is saved. Test results of the data acquisition system can fully meet the need of data acquisition in array data acquisition of lung sound.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.