Abstract

A small animal positron emission tomography (PET) instrument using a high-resolution solid-state detector insert in a conventional PET system was investigated for its potential to achieve sub-millimeter spatial resolution for mouse imaging. Monte Carlo simulations were used to estimate the effect of detector configurations (thickness, length and radius) on sensitivity. From this initial study, a PET system having an inner cylindrical silicon detector (4 cm ID, 4 cm length and 1.6 cm thickness composed of 16 layers of 300 µm × 300 µm × 1 mm pads), for scattering, surrounded by an outer cylindrical BGO scintillation detector (17.6 cm ID, 16 cm length and 2 cm thickness segmented into 3 mm × 3 mm × 20 mm crystals), for capture was evaluated in detail. In order to evaluate spatial resolution, sensitivity and image quality of the PET system, 2D images of multiple point and cylinder sources were reconstructed with the simulation data including blurring from positron range and annihilation photon acollinearity using filtered backprojection (FBP). Simulation results for 18F demonstrate 340 µm FWHM at the center of the field of view with 1.0% sensitivity from the coincidence of single scattering events in both silicon detectors and 1.0 mm FWHM with 9.0% sensitivity from the coincidence of single scattering in the silicon and full energy absorption of the second photon in the BGO detector.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.