Abstract
We report on the design, construction and current status of FOCS-2, the second continuous fountain microwave cesium frequency standard after FOCS-1. Both incorporate velocity-selective light traps driven by an electrostatic motor. FOCS-2 will take fuller advantage of the continuous fountain approach to gain in shot-noise-limited stability without loss of accuracy via the use of a higher flux. This is obtained via the implementation of a novel slow-atom pre-source and better collimation of the atomic beam. A detailed description of the apparatus is provided and compared with FOCS-1 to highlight improvements. In addition, we present results from related experiments on collimation in a 2D optical lattice. The goals for this new standard are a short-term stability of < 4 x 10-14 T -1/2 and a relative frequency uncertainty of < 1 x 10-5.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.