Abstract

Ionization chamber is considered as the golden standard for the dosemeter. This work fabricated one graphite-walled and two polymethyl methacrylate (PMMA)-walled ionization chambers, tested respectively and compared their characteristic parameters. The performance of graphite-walled chamber was similar to that of the PMMA-walled chamber. The relative energy response of ionization chambers was lower than 4% from 50keV to 50MeV based on the FLUKA code simulation results. The experimental data showed that the ionization chambers performed well in linearity and repeatability. Assuming the chambers sensitive volume and HV bias was constant, the measured maximum saturation absorbed dose of ionization chambers changed with the diameter variation of chambers anode based on the comparison results of two PMMA-walled ionization chambers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.