Abstract

Stereotactically placed guidewires are used for indicating the location of a nonpalpable carcinoma in breast-conserving surgery. Pathologists use the end of the embedded guidewire to guide sectioning during intraoperative margin assessment, but they do not currently have a tool to indicate the location of the guidewire end for informed sectioning. We present analysis and experimental testing of two optical methods for localizing the end of an embedded fiber-optic guidewire: the first uses irradiance emitted from the fiber to indicate the location of the guidewire end, while the second system uses the fiber optic to create a photoacoustic pulse for localization. Both systems locate the end of the guidewire within ±5 mm, which ensures that the lesion of interest is bisected during sectioning. The accuracy of the irradiance-based beacon is influenced by standard margin paints, so the photoacoustic beacon proved more useful under current tissue-handling protocols.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.