Abstract

In external fixation, bone screw loosening still presents a major clinical problem. For this study, the design factors influencing the mechanics of the bone-screw interface were analysed and various experimental screws designed with the intention of maximizing the strength and stiffness of the inserted screw. Push-in, pull-out and bending tests were then carried out on the three experimental screws, and on two commercially available screws in both a synthetic material and in cadaveric bone; photoelastic tests on different screw threadforms were also performed. The results of the push-in and pull-out tests indicate that both the screw threadform and cutting head have a significant effect on the holding strength of the screw. The photoelastic tests show that most of the applied load is distributed over the first few threads closest to the load, and that the area between the thread crests is subjected to high shear stresses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.