Abstract

A readout circuit for the linear array pH sensor was designed and developed, using chopper stabilization technique. It consisted of analog multiplexer, chopper stabilized operational amplifier, level shift, filter and buffer stage. The operational amplifier used in this design was a folded cascade amplifier. Results showed that the circuit can reduce input offset, drift voltage and noise. Moreover, noise can be further reduced on a faster chopper frequency. This readout circuit can be used for measuring small signal of low frequency biosensor. The readout circuit design was implemented using TSMC 0.35um 2P4M standard CMOS process.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.