Abstract

A multi-pulse Nd:YAG laser was designed and built, that can be used as an excitation source for Laser Induced Breakdown Spectroscopy (LIBS) experiments. A trigger and a power supply for the flash lamp has been successfully developed. A cooling system that uses a distilled water flow for the active medium was implemented. The laser has an output energy of 306 ± 7 mJ for 1 Hz repetition rate and its temporal profile is multi-pulse with 1 ps of average separation between them. To validate the system, the output was used as an ablation source of an aluminum target and the emission was investigated by spectroscopy analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.