Abstract

ABSTRACT This study distinguishes the design and analysis of a coaxial probe for measurement of biological body dielectric properties, in this measurement estimating the human tissue-equivalent liquid (TEL) permittivity and conductivity, to monitor and maintain the international standards for specific absorption rate (SAR) evaluation over the frequency band of 800 MHz-5 GHz. In addition, deionized (DI) water and ethanediol dielectric properties have been evaluated and the designed probe results compared to the commercial Dielectric Assessment kit (DAK) 3.5 probe. The obtained results are in good agreement with each other, moreover, the SAR calculation and each source of uncertainty budget analysis are estimated. Therefore, this fabricated probe may be suitable for liquid dielectric properties measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.