Abstract

Our objectives were to determine the validity and reproducibility of measurements on stereolithographic models and 3-dimensional digital dental models made with an intraoral scanner. Ten dry human skulls were scanned; from the scans, stereolithographic models and digital models were made. Two observers measured transversal distances, mesiodistal tooth widths, and arch segments on the skulls and the stereolithographic and digital models. All measurements were repeated 4 times. Arch length discrepancy and tooth size discrepancy were calculated. Statistical analysis was performed by using paired t tests. For the measurements on the stereolithographic and digital models, statistically significant differences were found. However, these differences were considered to be clinically insignificant. Digital models had fewer statistically significant differences and generally the smallest duplicate measurement errors compared with the stereolithographic models. Stereolithographic and digital models made with an intraoral scanner are a valid and reproducible method for measuring distances in a dentition.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.