Abstract

An attempt was made to develop and validate a simple, accurate high-performance thin-layer chromatographic (HPTLC) method for the simultaneous determination of diosmin, hesperidin, and ascorbic acid in pharmaceutical preparations. Analysis was performed on 10 cm × 20 cm aluminum-backed plates coated with 0.2 mm layers of silica gel 60 F254 (Merck, Germany). CAMAG TLC Scanner III was used for UV densitometric scanning. The used HPTLC system was found to give sharp, symmetrical, and well-resolved peaks at RF values of 0.34 ± 0.02, 0.40 ± 0.04, and 0.56 ± 0.03, and linearity was found in the ranges 100–800 ng/spot (r2 = 0.9985), 100–800 ng/spot (r2 = 0.9966), and 50–400 ng/spot (r2 = 0.9951) for diosmin, hesperidin, and ascorbic acid, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.