A novel approach was proposed to denoise quadrature Doppler signals from bi-directional blood flow using the wavelet frame and a soft-thresholding algorithm. A direction separation step was carried out first to avoid the phase distortion of quadrature Doppler signals, which is induced from the nonlinear, soft-thresholding processing. Then real parts of separated complex signals from the unidirectional flow were denoised independently. The quadrature Doppler signals from the bi-directional flow were reconstructed from the denoised separated signals. The approach has been applied to the simulated Doppler signals from a femoral artery. It is concluded from the experimental results that this method is practical for denoising quadrature Doppler signals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.