Abstract

Demoxepam, when derivatized by silylation and analyzed using gas chromatography-mass spectrometry (GC-MS), produces artifacts which are falsely identified as nordiazepam and oxazepam. Demoxepam was analyzed unextracted at various concentrations, using different derivatization procedures, and on different GC-MS systems. Oxazepam and nordiazepam were consistently identified in neat demoxepam samples, despite the changing variables. Under certain conditions, oxazepam was identified as low as 50 ng/mL derivatized demoxepam, and nordiazepam identified as low as 500 ng/mL derivatized demoxepam. The analysis of underivatized demoxepam resulted in nordiazepam detection at levels ≥2,500 ng/mL, whereas oxazepam was not detectable at or below 10,000 ng/mL demoxepam. Isolating the derivatization procedures and GC-MS analyses demonstrates that these processes are responsible for any degradation or rearrangement reactions which are taking place. Laboratories which follow similar procedures for benzodiazepine confirmations should consider these findings when interpreting analytical data from chlordiazepoxide cases.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.