Abstract

Assay development, assay validation, and documentation are reported here for a single packed column pressurized fluid chromatographic/ultraviolet (UV) method that provides: (1) simultaneous detection and quantification for the chiral drug, the chiral impurity and seven achiral impurities; and (2) a Fourier transform infrared (FT-IR) spectrometric identification test result for the Searle drug substance sample, xemilofiban. The separation is achieved in less than 30 min with three columns in tandem and a gradient of CO2-CH3OH. The post-column flow is split between UV (assay) and FT-IR (identification). Precision and accuracy are consistent within figures of merit obtained by liquid chromatographic-ultraviolet assays on analogous drug substances. The reported procedure combines three typical drug substance tests into one test (e.g. chiral impurities, achiral impurities, and infrared identification).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.