Abstract

High precision is required for ophthalmic robotic systems. This paper presents the kinematic calibration for the delta robot which is part of the next generation of Steady-Hand Eye Robot (SHER). A linear error model is derived based on geometric error parameters. Two experiments with different ranges of workspace are conducted with laser sensors measuring displacement. The error parameters are identified and applied in the kinematics to compensate for modeling error. To achieve better accuracy, Bernstein polynomials are adopted to fit the error residuals after compensation. After the kinematic calibration process, the error residuals of the delta robot are reduced to satisfy the clinical requirements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.