Abstract

For delayed coincidence experiments using NaI(Tl) scintillation counters, first photoelectron triggering (FPET) has been practicable since the advent of the RCA 8575. Using FPET, improved time resolution at low y ray energies was first demonstrated by Lynch, who also described his results by a model that fails to take into account the photomultiplier response. Using FPET, we have done systematic experiments from 6-511 keV for three different crystal sizes. We describe our results quantitatively by a simple model that takes into account the statistics of photoelectron production and the photomultiplier response, similar to El-Wahab and Kane. From the model calculations, we conclude that with present photomultipliers and NaI(Tl) scintillators, further improvements in resolution cannot be expected. The major difficulty is not the light output of NaI(Tl), but the single electron response of the photomultiplier.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.