Abstract
Allowable total error (ATE) are performance specification limits predefined for a variety of laboratory analytes. These limits define the maximum amount of error that is allowed for an assay when judging acceptability of a new assay during method verification/validation, evaluating patient or instrument comparison data, or in designing a quality control strategy. There are several widely available resources and models that can serve as a guide in selecting ATE. They may be based on legal requirements or set by providers of proficiency testing (PT) and external quality assessment schemes (EQAS). ATE can be also determined by professional expert groups or be based on biological variation of an analyte. Because there are several resources to choose from, there have been several attempts in reaching consensus on which ATE resource should be given preference. This chapter reviews several of these resources in more detail and discusses the difference between allowable total error (ATE) and observed total analytical error (TAE).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.