Abstract

A study has been made of the fractures and associated material from 34 prostheses broken in service. All the fractures appear to be of the fatigue type. Search by metallographic analysis for cracklike generating defects in material near the fracture reveals such conditions as abnormally coarse crystals, large nonmetallic inclusions, large inclusion population, undissolved master alloy particles, abnormal carbide segregation, interdendritic segregation, shrinkage, and gas porosity. Recommendations are made for the choice of metallic alloys for femoral stem application.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.