Abstract

To observe the influence of circulatory changes on the lag time of the pulse oximeter response, eight healthy patients scheduled for hand surgery were studied. After breath holding, the patients took a breath of oxygen and the time to an increase in SpO2 was measured before and after axillary brachial plexus block. It was found that the lag time with finger probe decreased from 28.6 +/- 7.1 sec to 15.8 +/- 1.1 sec (mean +/- SD) following brachial block (P less than 0.01). There was no change in arterial blood pressure. The results suggest that the lag time of the finger pulse oximeter response is primarily determined by blood flow.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.