Abstract
This study primarily focuses on improving the energy resolution of Double-Sided Silicon Strip Detectors (DSSSD) through optimized data analysis procedures using General Electronics for Time Projection Chambers (GET). We introduce two edge identification algorithms that demonstrate comparable efficacy. These methods are subsequently integrated with various techniques for measuring pulse amplitude, including sample smoothing and pole-zero cancellation. Notably, sample smoothing significantly enhances performances, achieving a resolution of 0.33% with DSSSD α calibration data. Additionally, we describe and apply the trapezoidal filter, examining its impact on resolution improvement and obtaining comparable results. The study also evaluates how internal GET parameters, specifically the sampling frequency and the shaping time of the Sallen-Key (SK) low-pass filter, affect signal resolution. Higher sampling frequencies and lower values of the SK filter were found to increase performance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.