Abstract

A humanoid thorax phantom containing six compartments was scanned with two different computed tomography (CT) scanners using various image acquisition and reconstruction parameters. The differences of CT numbers were statistically significant between the two CT scanners for each compartment ( p<0.001) except for the “air” compartment. The variabilities of the CT numbers are described for the different parameters. The mean CT numbers of the “water” compartment, for instance, ranged from 1 to 15 HU (Hounsfield Units), those of the “air” compartment varied from −962 to −990 HU. Knowledge of these CT number variabilities is necessary when CT numbers are used for tissue characterization.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.