Abstract

Abstract Purpose The aim of this study was to compare the accuracy of colonoscopy (CS) and CT colonography (CTC) in the measurement of colorectal polyps using pathological size as a reference. Materials and methods The analysis included 61 colorectal polyps in 28 patients who underwent preoperative CTC at our institution. All polyps were endoscopically resected. Polyp sizes were measured by CS and CTC. Endoscopic polyp size was extracted from endoscopy records written by one of two endoscopists (A with 11 and B with 6 years of endoscopic experience, respectively), who estimated the size visually/categorically without any measuring devices. After matching the location, the polyp size was measured on CTC using manual three-dimensional (3D) measurement on a workstation. The sizes of resected polyps were also measured after pathological inspection. Differences of the polyp size between CTC and histology, and between CS and histology were compared using paired t tests. Differences in measurement between the two endoscopists were also analyzed. Results The mean diameters of polyps measured using CS, CTC, and pathology were 10.5 mm, 9.2 mm, and 8.4 mm, respectively. There was a significant correlation between CS and pathology, as well as between CTC and pathology (both P < 0.0001). The correlation coefficient for CS (r = 0.86) was lower than that for CTC (r = 0.96). The correlations between CS and pathology for endoscopists A and B were 0.90 and 0.89, respectively. Conclusion Measurements of polyp size using CTC were closer to the pathological measurements compared to those by CS, which exhibited greater variability. This suggests that CTC may be more suitable for polyp size measurements in the clinical setting if patients undergo CTC concurrently with colonoscopy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call