To bridge the gap between traditional multilayer and crystal optics a high-resolution multilayer monochromator with a bandwidth of 0.22% has been designed and installed on a bending-magnet beamline (F3) at the Cornell High Energy Synchrotron Source (CHESS) to provide an unfocused monochromatic X-ray beam for protein crystallography experiments. Crystallographic data of excellent quality from a medium-sized protein, Concanavalin A, were collected and processed using standard crystallographic programs. The data were successfully used for a structure solution and refinement. The flux from the multilayer monochromator is enhanced, relative to that from a flat Si(111) monochromator, by a factor of 5; consequently, data collection is faster and/or smaller samples may be used. At the same time, the bandwidth is narrow enough to avoid streaked spots. This experiment suggests that multilayer optics may play a valuable role in satisfying the demands of the structural biology community for rapid X-ray data collection, particularly at under-utilized bending-magnet beamlines.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.