Abstract

A coupled LC-GC method for the analysis of carbazole-type PANH has been developed and evaluated. Group separation and isolation of carbazoles from interfering acridines in a complex sample matrix was accomplished by using a back-flush technique and an in situ end-capped dimethylaminopropyl silica column in the HPLC part of the system. On-line injection of the carbazole fraction into the GC column was performed with a loop-type interface utilizing concurrent solvent evaporation technique. An LOD of 1–3 pg of individual carbazole compounds was achieved by nitrogen selective detection using an NPD. The method is shown to be robust and is demonstrated by application to personal exposure measurement in an aluminum reduction plant.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.