Abstract

This study examines the application of coupled-column LC-photochemically induced fluorimetry-fluorescence detection (LC-LC-PIF-FD), demonstrating its potential for the quantitative and selective detection of six herbicides, including propanil and the phenylureas monuron, monolinuron, chlorotoluron, diuron and neburon in groundwater samples. An AQUASIL C18 50 x 4.6 mm(2) id column coupled to an AQUASIL C18 150 x 4.6 mm(2) id column for analyte clean-up and determination were used, respectively. A simple SPE with Cl8 cartridges was carried out, yielding average recoveries between 80 and 112% (n = 6) with RSDs between 0.5 and 9%. The LODs ranged from 0.0083 to 0.0833 microg/L in the groundwater samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.