Abstract

A coupled-column high-performance liquid chromatographic system for integrated, on-line sample processing and the determination of free and conjugated 1-hydroxypyrene in urine has been developed. The method is based on a “tailor-made” copper phthalocyanine-modified porous-glass precolumn packing material, which allows a direct and repeated injection of urine samples and a selective enrichment of trace amounts of particular components. The fully automated method has a low detection limit (0.01 pmol), a quantitative and matrix-independent recovery and a highly reliability, as shown by an interlaboratory comparison of methods.