Abstract

This paper reports a lab-on-a-chip device that counts the number of bacteria flowing through a microchannel. The bacteria number counting is realized by a microfluidic differential Resistive Pulse Sensor (RPS). By using a single microfluidic channel with two detecting arm channels placed at the two ends of the sensing section, the microfluidic differential RPS can achieve a high signal-to-noise ratio. This method is applied to detect and count bacteria in aqueous solution. The detected RPS signals amplitude for Pseudomonas aeruginosa ranges from 0.05 V to 0.17 V and the signal-to-noise ratio is 5–17. The number rate of the bacteria flowing through the sensing gate per minute is a linear function of the sample concentration. Using this experimentally obtained correlation curve, the concentration of bacteria in the sample solution can be evaluated within several minutes by measuring the number rate of the bacteria flowing through the sensing gate of this microfluidic differential RPS chip. The method described in this paper is simple and automatic, and have wide applications in determining the bacteria and cell concentrations for microbiological and other biological applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.