Abstract

A hybrid optical device that uses a multimode fiber coupled to a tunable light source for illumination and a 2.4-mm photodiode for detection in contact with the tissue surface is developed as a first step toward our goal of developing a cost-effective, miniature spectral imaging device to map tissue optical properties in vivo. This device coupled with an inverse Monte Carlo model of reflectance is demonstrated to accurately quantify tissue absorption and scattering in tissue-like turbid synthetic phantoms with a wide range of optical properties. The overall errors for quantifying the absorption and scattering coefficients are 6.0+/-5.6 and 6.1+/-4.7%, respectively. Compared with fiber-based detection, having the detector right at the tissue surface can significantly improve light collection efficiency, thus reducing the requirement for sophisticated detectors with high sensitivity, and this design can be easily expanded into a quantitative spectral imaging system for mapping tissue optical properties in vivo.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.