Abstract

Ti-Ni alloy has been increasingly applied to medical and dental devices, such as coronary stents and orthodontic wires. This alloy contains nickel, which is known to give rise to cytotoxicity, metal allergy, and carcinogenicity. Therefore, the purpose of this study was to improve the corrosion resistance of Ti-Ni alloy by electrolytic treatment, whereby investigation was carried out using different acidic electrolyte compositions. As a result, specimens electrolyzed with lactic acid, water, and glycerol were found to show higher corrosion potential and release lower amount of titanium and nickel ions than mechanical-polished specimens (p<0.05). With the electrolytic treatment, nickel concentration in the surface oxide layer of Ti-Ni alloy decreased, and the thickness of the surface oxide layer increased. Based on the results of this study, it was shown that electrolytic treatment with suitable electrolyte could improve the corrosion resistance of Ti-Ni alloy, which is effective to produce medical and dental devices that utilize shape memory effect or superelasticity with better biocompatibility.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.