Abstract

In this work, fracture behaviors of MRI compatible Nb-60Ta-2Zr alloy were investigated under quasi-static and fatigue loadings. The tensile fracture started with the decohesion of grain boundaries, and followed by microvoid coalescence. Fatigue tests were performed at ambient air as well as in simulated physiological solution with pH = 5.0 at R = 0.1. No deterioration happened with fatigue limit of 300 MPa when loaded in the corrosive solution. For both environments, the fatigue crack initiated at grain boundary and then propagated transgranularly. The grain boundary strength plays a key role in the fracture under both quasi-static and cyclic loadings.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.