Abstract

Optical coherence tomography (OCT) is a high-resolution three-dimensional imaging technique that enables nondestructive measurements of surface and subsurface microstructures. Recent developments of OCT operating in the mid-infrared (MIR) range (around 4 µm) lifted fundamental scattering limitations and initiated applied material research in formerly inaccessible fields. The MIR spectral region, however, is also of great interest for spectroscopy and hyperspectral imaging, which allow highly selective and sensitive chemical studies of materials. In this contribution, we introduce an OCT system (dual-band, central wavelengths of 2 µm and 4 µm) combined with MIR spectroscopy that is implemented as a raster scanning chemical imaging modality. The fully integrated and cost-effective optical instrument is based on a single supercontinuum laser source (emission spectrum spanning from 1.1 µm to 4.4 µm). Capabilities of the in situ correlative measurements are experimentally demonstrated by obtaining complex multidimensional material data, comprising morphological and chemical information, from a multilayered composite ceramic-polymer specimen.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.