Abstract

Zirconia implants that were restored with veneered zirconia displayed severe chipping rates of the restorations in clinical studies. Purpose of this study was to evaluate the fracture load of different zirconia implant supported monolithic crown materials (zirconia, alumina, lithium disilicate, feldspar ceramic and polymer-infiltrated ceramic) cemented with various cements (Harvard LuteCem SE, Harvard Implant Semi-permanent, Multilink Automix, VITA Adiva F-Cem). Flexural strength and fracture toughness of crown materials and compressive strength of the cements were measured. Fracture load values of crowns fabricated from lithium disilicate, feldspar ceramic and polymer-infiltrated ceramic were increased when cement with high compressive strength was used. Fracture loads for zirconia and alumina crowns were not influenced by the cement. Flexural strength and fracture toughness of the ceramics correlated linearly with the respective fracture load when using adhesive cement with high compressive strength. To achieve sufficient fracture load values, cementation with adhesive cement is essential for feldspar and polymer-infiltrated ceramic.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.