Abstract

In the IAEA-AAPM dosimetry formalism, detector measurements in general nonstandard conditions are corrected using the factor . This factor needs to be evaluated on a case-by-case basis which is difficult to accomplish in practice. The present paper aims to provide a method that allows neglecting correction factors for small and composite IMRT fields by first determining a radiation detector’s usability in these fields. Detailed models of nine radiation detectors are built: four ionization chambers (NE2571, A12, A1SL, A14), three small field detectors (PTW31018 microLion, PTW60003 natural diamond, PTW60012 unshielded diode) and two near water-equivalent detectors (alanine, W1 scintillating fiber). Using the egs_chamber Monte Carlo code, dose response functions at 6 MV and 25 MV are sampled for each detector and their corresponding volume of water. These functions are then used with a newly derived criterion to evaluate an upper bound on the variable if no field collimation/modulation occurs over a given perturbation zone. The variable is defined as the absolute value of the relative deviation from unity of a nonstandard field quality correction factor . Using the same criterion, perturbation zones are evaluated by finding the smallest field size allowed for correction-less dosimetry with a given tolerance . For composite fields, the sensitivity of detectors to the non-uniformity of virtual symmetric collapsed beams over regions of interest specified by the criterion is studied to estimate an upper bound on for a given beam flatness. Finally, a newly defined perturbation function is used to minimize the perturbations of the microLion chamber through density compensation. The theoretical criterion shows good agreement with full Monte Carlo simulations of . Perturbation zones are shown to be sensitive to both the energy of the beam and the orientation of the detector. The density-compensated microLion shows significantly improved response in both axial and radial orientations in small and composite IMRT fields. Finally, the new Exradin W1 scintillator is shown to have values under 1% in small fields. The methods presented in this work theoretically show that correction-less dosimetry of nonstandard field can be accomplished by knowing the limit of usability of radiation detectors in these conditions. Potential applications include small field output factor measurements and absolute absorbed dose to water verification in the QA of clinical IMRT fields.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.