Abstract

This paper investigates experimental underestimation of pressure measurements due to spatial averaging across a hydrophone sensitive element. Empirical relationships are measured to enable correction for hydrophone spatial averaging errors in peak compressional pressure (p c ), peak rarefactional pressure (p r ), and pulse intensity integral (pii). The empirical relationships show, for example, that if a 3-MHz, F/2 transducer is driven to moderate nonlinear distortion and measured at the focal point with a 500-μm membrane hydrophone, then spatial averaging errors are approximately 16% (p c ), 12% (p r ), and 24% (pii).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.