Abstract

One of the difficulties encountered in 19F NMR imaging of fluorinated blood substitutes is that these compounds often exhibit complex multipeak spectra. These peaks result in chemical-shift artifacts along the readout direction and blurred images. In addition, each peak excites a different slice (mis-selection) when a slice selection gradient is applied during the selective rf pulse. A simultaneous multislice imaging method has been developed to solve the inherent problem of mis-selection. The essence of this method is to use the two strongest peaks of the spectrum to excite controlled different multiple slices simultaneously, with or without a slice gap. The images corresponding to the two spectral lines are then separated from in- and out-of-phase images (Dixon method). This method corrects the problem of mis-selection and either improves the SNR or increases the number of slices over spectrally selective methods which image only one peak.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.