Abstract

Variable-rate selective excitation RF pulses modulate the slice selection gradients during RF transmission, especially to reduce the total RF power. Amplitude-modulated slice selection gradients can lead to increased gradient noise, in particular in high-field MRI where variable-rate selective excitation techniques are often used. In this work, an algorithm is presented that calculates a variable-rate selective excitation pulse modulation from given RF pulses with constant slice selection gradient. The algorithm avoids the known acoustic resonance frequencies of the gradient system to minimize sound pressure levels. It was tested with four different slice-selective RF pulse shapes (Sinc, Gaussian, and two Shinnar-LeRoux). Sound measurements revealed a reduction of the mean sound pressure level by up to 13 dB, and simultaneously, the specific absorption rate was reduced by 55%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.