Abstract

An efficient sampling method for dioxins from flue gas, an efficient automatic purification and fractionation method by a new HPLC system using a sulfuric acid–silica/silica column, a Nitro column and a PGC column, and sensitive determination method by an economical HRGC/LRMS using a programmable temperature vaporization (PTV) injector were developed. It was confirmed that the dioxins could be easily collected by the proposed simple sampling train consisting of only two bottles, and the extracted sample could be purified sufficiently and separated as mono-ortho PCBs, non-ortho PCBs and PCDDs/PCDFs completely with the proposed HPLC system. The peak areas of GC/MS were increased linearly with the injection volume up to 40 μl by the PTV technique, and the sensitivity could be increased to 20 times higher than usual. This convenient measuring method can drastically reduce operation time, consumption of hazardous solvent and cost.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.