Abstract

A very precise dosimetry of a field of artificial irradiation is essential to estimate an accurate ESR Equivalent Dose (ED) in archaeology and geology. Archaeological powder samples and one sample of DL-alanine powder as control were prepared by IPH. Some alanine dosimeters of LMRI, constituted of L-alanine pellets, were joined to the IPH samples and together, irradiated in boxes using panoramic 60Co source under different experimental conditions. The LMRI had in charge the control of the process of irradiation and the measurement of each absorbed dose and dose rate delivered to the different series of boxes. This study consists in the evaluation of the distribution of dose on the samples. The ESR measurements made by the two laboratories were analysed and then compared. These experiments showed a significant gradient in the absorbed dose delivered to each LMRI dosimeter. The dispersion in each box was easily observable. However, a good agreement between the dosimeters and the alanine powder was observed in a same box.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.