Abstract

The objectives were to determine whether the images obtained from different cone-beam computerized tomography (CBCT) field of view (FOV) sizes with varying scan time and kilovolt peak using the Picasso Master 3D machine (Vatech, Hwasung, South Korea) differ in contrast-to-noise ratio (CNR). A phantom was used. Scans were acquired using the same voxel size, 2 scan times (15 and 24 seconds) for 3 different FOV sizes and 5 different kilovolt peaks. The contrast and CNR were calculated for each FOV. The CNRs were similar for the 2 scan times for all 3 FOV sizes. The CNR was lower in the 20 × 19 cm FOV compared with the other 2 FOVs for both scan times. The CNR does not vary significantly with increasing scan times for all FOV sizes. Smaller FOVs result in enhanced CNR and contrast.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.