Abstract

To determine the effect of field of view (FOV) size on gray values in cone beam computed tomography (CBCT) and multislice computed tomography (MSCT) images. A phantom made up of 3 cylinders containing distilled water, plaster, and motor oil was constructed and inserted into an acrylic cylinder filled with distilled water. The phantom was scanned with a CBCT and MSCT device using 3 FOV sizes. Gray value of each material was evaluated in 40 axial slices, and the comparison between the results obtained with the same FOV size was made. In CBCT examinations, there was significant difference between the gray values of different FOVs for the 3 materials. In the MSCT, there was significant difference for the oil. Gray values showed significant difference between the CBCT and MSCT examinations for the 3 materials in the 3 different FOV sizes. Gray values determined in CBCT images are significantly influenced by the FOV size. Although the gray values obtained in MSCT have shown statistically significant differences between some acquisitions, the analysis of those differences seems to indicate low clinical relevance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.