Abstract

Purpose: To optimise breath-hold contrast-enhanced MR angiography (MRA) in infants and children with suspected congenital heart or thoracic vessel malformation. Material and Methods: Thirty-nine children (median age 1 year) were examined, using five different ultrafast MRA sequences with a TR between 3.2 and 5.0 ms and the contrast agent meglumine gadoterate. A test injection was used to determine contrast travel time. Different parameters for contrast injection were evaluated. Signal-to-noise ratio (SNR) measurements were performed and image quality and injection timing were evaluated. Results: MRA was successful in all patients and image quality was considered very good in 52%. Adequate SNR was achieved with no significant differences between the MR sequences. SNR decreased only 25-30% between subsequent scans. The mean contrast dose was 0.23 mmol/kg. The mean scan time was 12.5±3.8 s; the shorter scan times made dynamic examinations possible with high temporal resolution. Highest spatial resolution was obtained with TR 4.6/5.0 sequences. Conclusion: A contrast dose of 0.2 mmol/kg b.w. is recommended with an injection rate of 0.5 to 1.2 ml/s, depending on patient size and scan time. The scan delay time should equal the contrast travel time for optimal vessel enhancement. In the future, contrast-enhanced MRA may be a potential alternative to angiocardiography in infants and children.