Abstract

To investigate the agreement between continuous noninvasive blood pressure measurement with the ClearSight system (cNIBP-CS) and standard intermittent noninvasive blood pressure measurement (iNIBP) in patients with peripheral arterial disease (PAD). Additionally, the influence of vasoactive medication on potential measurement differences was assessed. A secondary analysis of a randomized controlled trial. At a university hospital. Thirty-four patients with PAD undergoing percutaneous transluminal angioplasty of the lower limbs. None. Continuous noninvasive blood pressures were measured with the "ClearSight" system and compared to standard iNIBPs. Bland-Altman analysis revealed a mean bias of 13 mmHg (±15) between cNIBP-CS and iNIBP, with 95% limits of agreement (LOA) ranging from -17 to 42 mmHg. When comparing both medication groups, a similar mean bias was found for phenylephrine and dobutamine (12 mmHg [±13] and 13 mmHg [±13], respectively). In this study, in patients with PAD, cNIBP-CS showed an underestimation of blood pressure compared to iNIBP in phenylephrine- and dobutamine-treated patients. Compared to previous studies, a larger bias and wider 95% LOA were found.