Abstract

Diode laser energy has been applied through a fiberoptic probe using a power setting of 2.5 watts (W) in the continuous mode. In this study we employed high-power diode laser energy (4 to 12 W, continuous wave) to incise ocular tissue through a fiberoptic probe using 100 microns and 300 microns tips. The retina was photocoagulated with a 300 microns orb tip. No bleeding occurred at the incision sites. Histologic evaluation revealed coagulation into the healthy tissue ranging from 10 to 50 microns.