Abstract
An efficient experimental system for the chemical reactions of various aqueous solutions using soft x rays from synchrotron radiation has been designed, constructed, and evaluated. This system was installed at an end station on a beamline (BL-11) at the Synchrotron Radiation Center of Ritsumeikan University. To protect the storage ring and the beamline from accidental vacuum breaks and contamination caused by unexpected solution leakage in the reaction apparatus, the system has two main protection sections: a differential vacuum pumping section equipped with orifice gaskets of very small aperture and an emergency interlocking system section equipped with a fast closing shutter, etc., in addition to a safety mechanism in the reaction apparatus section. The outline of the system and preliminary experimental data obtained in the evaluation of the system are described.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.