Abstract

We have developed a miniature electron ion coincidence (EICO) analyzer mounted on a conflat flange with an outer diameter of 114 mm. It consists of a cylindrical mirror analyzer (CMA), a time-of-flight ion mass spectrometer (TOF-MS), a commercially available linear motion feed through, and a tilt adjustment mechanism. Each sample surface was irradiated by synchrotron radiation, and the energies of emitted electrons were analyzed and detected by the CMA, while desorbed ions were collected by the TOF-MS in coincidence with the electrons. The performance of the EICO analyzer was tested by measuring the Auger-electron H+ photoion coincidence spectrum of condensed water at 4a1 <-- O 1s resonance.